ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Catalog Number SGC0705
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Thrombosis/Thrombus (4440); Unspecified Tissue Injury (4559)
Event Date 03/07/2022
Event Type  Injury  
Manufacturer Narrative
The device will not be returned for evaluation; the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is being filed to report an unspecified tissue injury.This will be filed.When the steerable guide catheter (sgc) was inserted into the left atrium, something similar to tissue was confirmed by echo on the tip.Upon removal of the sgc from the body, there was no deposit on the tip, and it was not possible to distinguish between thrombus and septal tissue.Additional aspiration was required, which is medical intervention to prevent a serious injury.There was no clinically significant delay.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the available information, the cause of the reported thrombus or tissue injury cannot be determined.The reported thrombus or tissue injury as listed in the instructions for use (ifu) are known possible complications associated with mitraclip procedures.The reported medical intervention was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13950648
MDR Text Key288202018
Report Number2024168-2022-03316
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/01/2022
Device Catalogue NumberSGC0705
Device Lot Number10902R124
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2022
Initial Date FDA Received03/30/2022
Supplement Dates Manufacturer Received04/27/2022
Supplement Dates FDA Received05/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight51 KG
-
-