• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. UNKN ENT REFLEX ULTRA PLASMA WAND AND TURBINATOR WAND; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. UNKN ENT REFLEX ULTRA PLASMA WAND AND TURBINATOR WAND; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number UNKNOWN
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Infection (1930)
Event Date 01/01/2006
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).Article cite: shao juan.(2006).Clinical observation of radiofrequency ablation in the treatment of obstructive sleep apnea hypopnea syndrome.Traffic medicine, 20(3), 328-329.
 
Event Description
It was reported that on literature review clinical observation of radiofrequency ablation for obstructive sleep apnea hypopnea syndrome, 3 patients had a local wound infection after a radiofrequency ablation procedure using a disposable reflex 55 coblation wand.The events were treated with antiinflammatory treatment which cause an improvement.Patients outcome is unknown.No further information is available.
 
Manufacturer Narrative
H10: h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record, complaint history, device labeling/ifu, risk management and capa, nc, pra/hhe review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.Without supporting clinical/medical documents, a thorough investigation cannot be performed.The root cause and/or patient outcome beyond that which was documented in the article could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.Should any additional clinical information be provided this complaint will be re-evaluated.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN ENT REFLEX ULTRA PLASMA WAND AND TURBINATOR WAND
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13954551
MDR Text Key288276223
Report Number3006524618-2022-00123
Device Sequence Number1
Product Code GEI
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Study,Literature,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2022
Initial Date FDA Received03/30/2022
Supplement Dates Manufacturer Received09/21/2022
Supplement Dates FDA Received09/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-