Catalog Number 9211101190
Device Problems Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005); Non Reproducible Results (4029)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2022
Event Type  malfunction  
Manufacturer Narrative
The most likely cause for the discrepancy observed may be sample specific and from lab cross-contamination.Also of note is the off-label collection kit in use.Throughout the data analysis, a systematic issue was not observed and a product problem was not found.Facility name: hong kong adventist hospital-(b)(6) rd 7/f,clinical laboratory, the facility's full name has been truncated due to character limitations.(b)(4).
 
Event Description
In light of the covid-19 pandemic and the subsequent emergency use authorizations (euas) for sars-cov-2 diagnostic tests, the agency has requested heightened reporting beyond the reasonably suggests requirements of 803 to include allegations of false positive or false negative results independent of harm or malfunction or off-label use.Pursuant to the agency¿s instruction, we hereby submit this mdr.A customer from hong kong alleged a discrepant result for a single patient while using the cobas® sars-cov-2 & influenza a/b nucleic acid test on the cobas® liat® system.The alleged sample initially generated a positive result for sars-cov-2.The same sample was retested on a different cobas® liat® system and a different platform (genexpert).Both of those tests yielded negative results.The initial positive result was not released.No harm was alleged.An investigation was conducted to evaluate the customer issue.Per fda¿s eua guidance, 1 mdr will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS® SARS-COV-2 & INFLUENZA A/B
Type of Device
COVID-19 MULTI-ANALYTE RESPIRATORY PANEL NUCLEIC ACID DEVICES
Manufacturer (Section D)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
branchburg NJ 08876
Manufacturer (Section G)
ROCHE MOLECULAR SYSTEMS, INC. - BRANCHBURG
1080 us highway 202 south
na
branchburg NJ 08876
Manufacturer Contact
stacie-ann creighton
1080 us hwy 202 s
na
branchburg, NJ 08876
9082537112
MDR Report Key13956077
MDR Text Key298051068
Report Number2243471-2022-00341
Device Sequence Number1
Product Code QLT
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
EUA201779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Catalogue Number9211101190
Device Lot Number11018Z
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/02/2022
Initial Date FDA Received03/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
-
-