Model Number 3851
Device Problem Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2022
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on (b)(6) 2022.It was reported that the device could not cross the lesion.A 10mmx3.00mm wolverine coronary cutting balloon was selected for use.During the procedure, it was noted that the device could not cross the lesion.The device was simply pulled out from the patient's body.The procedure was completed with a different device.No complications were reported and patient was good post procedure.However, device analysis revealed that a balloon pinhole was noted on the distal edge of the proximal marker band.
 
Manufacturer Narrative
Initial reporter facility name - national hospital organization osaka national hospital.Device evaluated by manufacturer: the device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was identified within the balloon which was evidence of a device leak.The returned device was attached to an encore inflation unit and positive pressure was applied in an attempt to inflate the balloon.Liquid was observed to be leaking from a balloon pinhole located at the distal edge of the proximal marker band.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.All blades were intact within their pads and fully bonded to the balloon material.A visual and tactile examination identified no issues with the hypotube or shaft polymer extrusion of this device.The markerbands and tip end of the device were visually and microscopically examined, and no issues were noted with the device that may have potentially contributed to the complaint incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13956402
MDR Text Key288366337
Report Number2134265-2022-03005
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/09/2023
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0028173001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/09/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-