Model Number CI-1600-05 |
Device Problems
Mechanical Problem (1384); Migration or Expulsion of Device (1395); Appropriate Term/Code Not Available (3191); Migration (4003)
|
Patient Problems
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 03/09/2022 |
Event Type
Injury
|
Event Description
|
The recipient reportedly is a non user due to lack of benefit.Programming adjustments were made, however, the issue did not resolve.A ct scan confirmed multiple electrodes outside of the cochlea.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The recipient experienced electrode migration prior to revision surgery.The external visual inspection revealed tool damage in the silicone overmold on the top cover, as well as cuts on the bottom cover and fantail region.In addition, the electrode was severed.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed some of the electrical tests performed.The device passed the mechanical test performed.This device was explanted for medical reasons.However, this device had a short in the electrode pocket.A corrective action was implemented.This version of the ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|