SMITH & NEPHEW, INC. R3 OFFSET IMPACTOR TIP; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 71363077 |
Device Problem
Material Fragmentation (1261)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/09/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Internal complaint reference: case (b)(4).
|
|
Event Description
|
It was reported that, during total hip replacement, an r3 offset impactor tip cracked in pieces inside the patient, when putting the shell in.No pieces fell into the patient.Surgery was resumed, without any delay, with a smith & nephew back-up device.Patient was not injured as consequence of this problem.
|
|
Manufacturer Narrative
|
H10: the device was not returned for evaluation and the reported event could not be confirmed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Internal complaint reference: case-2022-00095877-1.
|
|
Search Alerts/Recalls
|
|
|