Model Number 174317 |
Device Problems
Detachment of Device or Device Component (2907); Mechanics Altered (2984); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Abdominal Pain (1685); Cellulitis (1768); Purulent Discharge (1812); Emotional Changes (1831); Erythema (1840); Hemorrhage/Bleeding (1888); Muscle Weakness (1967); Nausea (1970); Pain (1994); Abnormal Vaginal Discharge (2123); Dizziness (2194); Anxiety (2328); Discomfort (2330); Sleep Dysfunction (2517); Hematuria (2558); Dysuria (2684); Fluid Discharge (2686); Foreign Body In Patient (2687); Constipation (3274); Syncope/Fainting (4411); Cramp(s) /Muscle Spasm(s) (4521); Unspecified Tissue Injury (4559); Unintended Radiation Exposure (4565); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 01/25/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, after approximately a year after the laparoscopic surgery, the patient complained of pain.The hospital did an x-ray and found a small metal pin in the patient's abdominal cavity.The pin was removed from the patient.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, after approximately a year after the total laparoscopic hysterectomy, the patient complained of pain.The hospital did an x-ray and found a small metal pin in the patient's abdominal cavity.The pin was removed from the patient.The same piece was found when another device was checked.
|
|
Manufacturer Narrative
|
Additional information: removed a1, g3.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, approximately a year after the total laparoscopic hysterectomy where an endoscopic grasper was used, the patient complained of pain.The hospital did a computed tomography (ct) scan and found a small metal pin in the patient's pelvic cavity.The 5mm pin was removed from the patient.The same piece was found when another endoscopic grasper was checked.
|
|
Manufacturer Narrative
|
Additional information: b3, b5, g3.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, 224 days post-operative of a total laparoscopic hysterectomy where an endoscopic grasper was used, the pa tient complained of right side pelvic and right lower back pain.Due to the symptoms, the hospital performed an x-ray and found a small metal pin in the patient's pelvic cavity.A computed tomography (ct) scan was done after 7 days which indicated that the device was extraluminal.37 days later, the patient underwent a surgical procedure to remove the 5mm pin.The same piece was found when another endoscopic grasper was checked.
|
|
Manufacturer Narrative
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, photos represented a sample i nstrument was available for evaluation.Visual inspection noted that the inner shaft had disengaged from retraction slider, there was no sign of deformation or breakage.The knob component was inspected using a microscope and no indentation around the knob hole was observed.Functional test was precluded since reported instrument was not received.It was reported that a component disengaged from the device into the surgical cavity.The reported issue was confirmed based on the photo.The most likely cause was determined to be manufacturing related.Internal process improvements have been initiated to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
During a regular internal review of complaints, it was identified that some medwatch reports associated with an fca were submitted without a correction/removal number in field h9 of 3500a form.Medtronic have initiated capa pr 576617 to address this gap.As part of correction activities, this supplemental medwatch is being submitted to correct this issue and provide the appropriate correction/removal number in field h9.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The patient¿s attorney alleged a deficiency against the device.The product was used for a total laparoscopic hysterectomy where an endoscopic grasper was used.224 days post-operatively, the patient complained of right side pelvic and right lower back pain.Due to the symptoms, the hospital performed an x-ray and found a small metal pin in the patient's pelvic cavity.A computed tomography (ct) scan was done after 7 days which indicated that the device was extraluminal.37 days later, the patient underwent a surgical procedure to remove the 5mm pin.The same piece was found when another endoscopic grasper was checked.It was reported that after use, the patient experienced a retained screw broken off of device, pain, failure of device, abdominal and pelvic pain, difficulty defecating, sharp spasms, difficulty breathing and presyncope, suffering, loss of quality of life, fear and worry, fainting, vaginal pain, vaginal discharge, vaginal bleeding, dysuria, oozing pus, postprocedural wound cellulitis, weakness, lightheadedness, anxiety, interrupted sleep, foreign body, blood in urine, difficulty urinating, constipation, discolored discharge, redness and discoloration, nausea, discomfort, lower back pain, and nerve pain.Post-operative patient treatment included revision surgery, removal of retained pin, medications, x-ray, and ct scan.
|
|
Manufacturer Narrative
|
Additional info: d10 (rumi ii koh 3 cm / endo 34cm surgiflo sterile applicator).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|