• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS IMAGE GUIDED THERAPY CORPORATION ANGIOSCULPT EVO RX PTCA; CATHETER, PTCA, CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS IMAGE GUIDED THERAPY CORPORATION ANGIOSCULPT EVO RX PTCA; CATHETER, PTCA, CUTTING/SCORING Back to Search Results
Model Number 2200-2020-B
Device Problem Entrapment of Device (1212)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2022
Event Type  Injury  
Manufacturer Narrative
The patient's dob or age at time of event, gender, weight, ethnicity, and race are unknown.This information was not available from the facility.Patient information regarding relevant tests/laboratory data or medical history are unknown.This information was not available from the facility.Foreign: (b)(6).The angiosculpt device was discarded by the facility, thus no returned product investigation was performed.Per the ifu, if unusual resistance is felt when the catheter is being manipulated, carefully remove the entire catheter system.
 
Event Description
The angiosculpt device was used in a moderately tortuous and moderately calcified rca.After inflation/deflation and upon removal, the angiosculpt got stuck on the guide wire, thus were removed as a unit.Outside of the patient, the guidewire was dislodged from the angiosculpt and reinserted into the patient.The procedure was completed with a non-angiosculpt device.No patient injury reported.This adverse event is being reported because the angiosculpt got stuck on a non-manufacturer's guide wire and removed as a unit.Placement of a new guide wire was necessary to complete the procedure.
 
Manufacturer Narrative
Blocks d1/ g1: manufacturer name and site name were updated from spectranetics to philips image guided therapy corporation to match the information on the product label.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOSCULPT EVO RX PTCA
Type of Device
CATHETER, PTCA, CUTTING/SCORING
Manufacturer (Section D)
PHILIPS IMAGE GUIDED THERAPY CORPORATION
5055 brandin court
fremont CA 94538
Manufacturer (Section G)
PHILIPS IMAGE GUIDED THERAPY CORPORATION
5055 brandin court
fremont CA 94538
Manufacturer Contact
ana tan
5055 brandin court
fremont, CA 94538
MDR Report Key13970897
MDR Text Key289608336
Report Number3005462046-2022-00016
Device Sequence Number1
Product Code NWX
UDI-Device Identifier00813132026820
UDI-Public00813132026820
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P050018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number2200-2020-B
Device Catalogue Number2200-2020-B
Device Lot NumberG21120005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNK MFG AND SIZE: GUIDE CATHETER; UNK MFG AND SIZE: GUIDE WIRE; UNK MFG AND SIZE: INTRODUCER SHEATH
Patient Outcome(s) Other;
-
-