Product analysis: it was determined on analysis that the programmer tested out of specification as it was identified that the stylus was defective.It was also determined at analysis that the floppy drive was defective.The display left locking latch hasp was worn and the left and right locking latch of the cord bay door was broken.The upper and lower bullets were worn, and the left keyboard hinge was broken.The cooling fan was noisy and the electrocardiogram (ecg).Errors were found in the software history.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|