Model Number 186715000
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/04/2022
Event Type  Injury  
Manufacturer Narrative
Additional narrative: additional device product codes: kwp; kwq; mnh; mni; osh.Complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2022 during spinal fusion procedure set screw stripped and potentially left metal shards in the patient.Entire screw and set screw were replaced.Surgery was completed with a delay of thirty (30) minutes.Patient outcome is unknown.This report is for one (1) mis single inner setscw.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: a manufacturing record evaluation was performed for the finished device.Product code: 186715000, lot: 322413.Was electronically reviewed and no non-conformances / manufacturing irregularities were identified during the manufacturing process.The product was released on: september 23, 2021.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the mis single inner setscw was found with broken threads and jammed to the wide blade hook.No other anomalies were noted.The embedded device condition cannot be confirmed since x-ray evidence was not provided.A dimensional inspection was unable to be performed due to the post-manufacturing damage and returned condition.A functional test was not performed as the mating device was not returned.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the mis single inner setscw would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: current and manufactured.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIS SINGLE INNER SETSCW
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
8472871282
MDR Report Key13978037
MDR Text Key288389159
Report Number1526439-2022-00518
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034121192
UDI-Public(01)10705034121192
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number186715000
Device Catalogue Number186715000
Device Lot Number322413
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
VIPER UNI SCREW 6X40MM TI; VIPER UNI SCREW 6X40MM TI; WIDE BLADE HOOK
Patient Outcome(s) Required Intervention;
Patient Age14 YR
Patient SexFemale
-
-