MEDTRONIC HEART VALVES DIVISION AVALUS AORTIC TISSUE VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 40027
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problem Insufficient Information (4580)
Event Date 03/03/2022
Event Type  Injury  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that the replica end of the sizing tool for this 27mm aortic bioprosthetic valve were used during the sizing process, the surgeon chose to attempt the 27mm valve for a greater effective orifice area (eoa) and a body surface area (bsa) chart was not used for valve sizing.It was reported that after pledget placement and suturing, the surgeon decided this 27mm bioprosthetic valve was too large, the valve was explanted and replaced with a 25mm valve of the same model.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVALUS AORTIC TISSUE VALVE
Type of Device
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key13979737
MDR Text Key289018341
Report Number2025587-2022-00872
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P170006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/05/2024
Device Model Number40027
Device Catalogue Number40027
Was Device Available for Evaluation? No
Date Manufacturer Received03/04/2022
Date Device Manufactured04/06/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-