Model Number 1458Q/86 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Further information was requested but not received.
|
|
Event Description
|
It was reported that the patient presented for lead implant procedure.During the procedure, the left ventricle lead to be implanted exhibited detachment of device or device component.The physician did not use the lead during the procedure on (b)(6) 2022.The patient was stable.
|
|
Event Description
|
It was reported that the patient presented for lead implant procedure.During the procedure, the left ventricle lead to be implanted exhibited detachment of device or device component and the issue was discovered while the left ventricle lead was being pulled out of the vein and was in cusp of coronary sinus.The physician felt the tip of the left ventricle lead was loose during repositioning.The left ventricle lead was replaced during the procedure on (b)(6) 2022.The patient was stable.
|
|
Manufacturer Narrative
|
The reported event of ¿lv lead tip disconnection with the lead body¿ was confirmed.As received, a complete lead was returned in one piece with stylet stuck inside the lead.The ptfe coating of the stylet was found stripped and clogged in the inner coil distal to the connector pin.The connector pin and connector cap were found pulled out of the connector assembly and the inner coil was stretched consistent with excessive forces during the procedure.The cause of the reported event was isolated to procedural damage.
|
|