• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Further information was requested but not received.
 
Event Description
It was reported that the patient presented for lead implant procedure.During the procedure, the left ventricle lead to be implanted exhibited detachment of device or device component.The physician did not use the lead during the procedure on (b)(6) 2022.The patient was stable.
 
Event Description
It was reported that the patient presented for lead implant procedure.During the procedure, the left ventricle lead to be implanted exhibited detachment of device or device component and the issue was discovered while the left ventricle lead was being pulled out of the vein and was in cusp of coronary sinus.The physician felt the tip of the left ventricle lead was loose during repositioning.The left ventricle lead was replaced during the procedure on (b)(6) 2022.The patient was stable.
 
Manufacturer Narrative
The reported event of ¿lv lead tip disconnection with the lead body¿ was confirmed.As received, a complete lead was returned in one piece with stylet stuck inside the lead.The ptfe coating of the stylet was found stripped and clogged in the inner coil distal to the connector pin.The connector pin and connector cap were found pulled out of the connector assembly and the inner coil was stretched consistent with excessive forces during the procedure.The cause of the reported event was isolated to procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13996646
MDR Text Key288502424
Report Number2017865-2022-06970
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000116014
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/24/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-