• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. OPEN IMPLANT CLOSURE TOP, 5.5MM, TI; SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. OPEN IMPLANT CLOSURE TOP, 5.5MM, TI; SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM Back to Search Results
Catalog Number 3301-1
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2022
Event Type  malfunction  
Event Description
It was reported that three pathfinder closure tops stripped during final tightening intra-operatively.They were removed and replaced with other closure tops to complete the procedure without patient impacts.This is report two of three for this event.
 
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2022-00083 through 3012447612-2022-00085.
 
Event Description
It was reported that three pathfinder closure tops stripped during final tightening intra-operatively.They were removed and replaced with other closure tops to complete the procedure without patient impacts.This is report two of three for this event.
 
Manufacturer Narrative
Device evaluation: visual inspection revealed no damage to either the hex drive or the threads.Potential cause: root cause was unable to be determined as no product issue was found.Dhr review: per dhr review, the parts were likely conforming when they left zimvie control.Device use: these devices are used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN IMPLANT CLOSURE TOP, 5.5MM, TI
Type of Device
SEQUOIA PEDICLE SCREW SYSTEM INCLUDING SPEEDLINK II TRANSVERSE CONNECTOR SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key13996699
MDR Text Key288855733
Report Number3012447612-2022-00084
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00889024334861
UDI-Public(01)00889024334861(10)ABD
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K131980
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3301-1
Device Lot NumberABD
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/11/2022
Initial Date FDA Received04/04/2022
Supplement Dates Manufacturer Received06/07/2022
Supplement Dates FDA Received06/27/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/22/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-