Catalog Number 3301-1 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/11/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that three pathfinder closure tops stripped during final tightening intra-operatively.They were removed and replaced with other closure tops to complete the procedure without patient impacts.This is report two of three for this event.
|
|
Manufacturer Narrative
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2022-00083 through 3012447612-2022-00085.
|
|
Event Description
|
It was reported that three pathfinder closure tops stripped during final tightening intra-operatively.They were removed and replaced with other closure tops to complete the procedure without patient impacts.This is report two of three for this event.
|
|
Manufacturer Narrative
|
Device evaluation: visual inspection revealed no damage to either the hex drive or the threads.Potential cause: root cause was unable to be determined as no product issue was found.Dhr review: per dhr review, the parts were likely conforming when they left zimvie control.Device use: these devices are used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
|