Model Number UHI-4
Device Problem Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2022
Event Type  malfunction  
Manufacturer Narrative
The referenced device was returned to olympus for service/repair.The evaluation found that after switching on the power source an alarm is generated and the light on the front panel turns off which is attributed to a defective (cr) board (pressure sensor/pressure sensor circuit).The cause of the defective board could not be confirmed.However, the investigation is ongoing; therefore, the root cause of the reported issue cannot be determined at this time.If additional information becomes available a follow up medical device report will be supplemented accordingly.
 
Event Description
The olympus (osh) field service engineer was informed that the customer high flow insufflation unit reportedly doesn¿t work¿ during reprocessing prior to a cholecystectomy procedure.The facility completed the procedure with another similar unit.No death, injury or harm was reported to olympus.
 
Manufacturer Narrative
This report is being supplemented to provide additional information, based on the approved final investigation.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the faulty circuit board could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14003053
MDR Text Key298210956
Report Number8010047-2022-05648
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/07/2022
Initial Date FDA Received04/04/2022
Supplement Dates Manufacturer Received04/27/2022
Supplement Dates FDA Received05/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-