Model Number UHI-4 |
Device Problem
Defective Device (2588)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The referenced device was returned to olympus for service/repair.The evaluation found that after switching on the power source an alarm is generated and the light on the front panel turns off which is attributed to a defective (cr) board (pressure sensor/pressure sensor circuit).The cause of the defective board could not be confirmed.However, the investigation is ongoing; therefore, the root cause of the reported issue cannot be determined at this time.If additional information becomes available a follow up medical device report will be supplemented accordingly.
|
|
Event Description
|
The olympus (osh) field service engineer was informed that the customer high flow insufflation unit reportedly doesn¿t work¿ during reprocessing prior to a cholecystectomy procedure.The facility completed the procedure with another similar unit.No death, injury or harm was reported to olympus.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information, based on the approved final investigation.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the investigation, the definitive root cause of the faulty circuit board could not be determined.Olympus will continue to monitor field performance for this device.
|
|