• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD FEMORAL HEAD 48MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD FEMORAL HEAD 48MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Model Number 74123148
Device Problems Biocompatibility (2886); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Failure of Implant (1924); Metal Related Pathology (4530)
Event Date 12/22/2020
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Event Description
(b)(4), it was reported that, after a bhr resurfacing construct had been implanted on the plaintiff¿s right hip on (b)(6) 2012, the plaintiff experienced pain, aseptic loosening of the resurfacing head, and adverse local tissue reaction due to metallosis around the hip capsule.Revision surgery was performed on (b)(6) 2020 to treat this adverse event.During this procedure, the femoral head 48mm was explanted and replaced with a total hip arthroplasty construct.The patient was transferred to the recovery area in stable condition.
 
Manufacturer Narrative
H3, h6: it was reported that a right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment have not been returned for evaluation.A review of the historical complaints data for the acetabular cup and femoral head was performed using batch numbers to evaluate patterns of repeated failures or defects over the lifetime of the product and using part numbers and the reported failure modes to evaluate patterns of repeated failures or defects in a timeframe prior 12 months as of the complaint aware date.No other similar complaints were identified to involve this batch.Other similar complaints were identified for the part number and the reported failure mode.This will continue to be monitored.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions required.The available medical documents were reviewed.The clinical information provided, of aseptic loosening of right hip resurfacing with adverse local tissue reaction may be consistent with a reaction to metal debris.However, the source and the root cause cannot be determined with the available documentation.It cannot be concluded that the reported clinical findings are associated with the implant failure.Based on the available information we can confirm the reported complaint, however without further information our investigation remains inconclusive, and a definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD 48MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14008084
MDR Text Key288590090
Report Number3005975929-2022-00119
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010552426
UDI-Public03596010552426
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/27/2016
Device Model Number74123148
Device Catalogue Number74123148
Device Lot Number11KW34770
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/09/2022
Date Device Manufactured11/29/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other; Hospitalization;
Patient Age60 YR
Patient SexFemale
-
-