MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10013
Device Problems Connection Problem (2900); Mechanics Altered (2984); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/22/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).The sample is under evaluation by manufacturing site.
 
Event Description
It was reported that, during a cori assisted tka surgery, the looking nuts of the real intelligence robotic drill were loose and could not attach the long attachment into the drill.For the second real intelligence robotic drill they were unable to load the burr into the drill and received a message saying: "the robotic drill has been deactivated press continue to reinitialize the robotic drill".Then, the third real intelligence robotic drill was working but slow, sometimes showing the error message "robotic drill cable disconnected".The procedure was finished with a smith and nephew back up device without delay.The patient was not harmed.For the third drill it was found during investigation that the exposure motor work intermittent, which makes this a reportable event.
 
Manufacturer Narrative
H3, h6: the real intelligence robotic drill, rob10013, sn (b)(6), used for treatment was returned for evaluation.A relationship between the reported event and the device was established.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported event was confirmed.At the start of the evaluation, the device failed testing and displayed critical error message.The device motor worked intermittently.The device was disassembled without issue, to inspect the components.Motor was removed and reinstalled and it operated as designed.During this evaluation, the device passed kpc testing and a case was reviewed successfully.The most likely cause of this event is an electrical communication failure of the drill exposure motor encoder.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.A historical escalation event review was completed.The review determined that prior escalation actions are applicable to the scope of this complaint and further investigation into the reported failure is being conducted to determine if additional escalation actions are required.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Further investigation into the reported failure is being conducted to determine if additional actions are required.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Manufacturer Narrative
H3, h6: the real intelligence robotic drill, rob10013, (b)(6), used for treatment was returned for evaluation.A relationship between the reported event and the device was established.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported event was confirmed.At the start of the evaluation, the device failed testing and displayed critical error message.The device motor worked intermittently.The device was disassembled without issue, to inspect the components.Motor was removed and reinstalled and it operated as designed.During this evaluation, the device passed kpc testing and a case was reviewed successfully.The most likely cause of this event is an electrical communication failure of the drill exposure motor encoder.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.A historical escalation event review was completed.The review determined that prior escalation actions are applicable to the scope of this complaint and further investigation into the reported failure is being conducted to determine if additional escalation actions are required.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Further investigation into the reported failure is being conducted to determine if additional actions are required.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Manufacturer Narrative
H7: if remedial action initiated, check type and h9: if action reported to fda under 21 usc 360i(f), list correction/removal reporting number.Section h3, h6: the real intelligence robotic drill, rob10013, (b)(6), used for treatment was returned for evaluation.A relationship between the reported event and the device was established.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.The reported event was confirmed.At the start of the evaluation, the device failed testing and displayed critical error message.The device motor worked intermittently.The device was disassembled without issue, to inspect the components.Motor was removed and reinstalled and it operated as designed.During this evaluation, the device passed kpc testing and a case was reviewed successfully.The most likely cause of this event is associated with a failure of the drill exposure motor due to the thermo-mechanical stress induced within the motor at the encoder and electrical noise on the console error status inputs to the drill exposure motor encoder.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found similar events.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A historical review concluded that the lot, serial number or part number reported in this event is related to a corrective/preventive action already implemented.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Continuous improvements have been made to the cori robotic drill and manufacturing processes to reduce drill disconnection error messages.These improvements consisted of: 1.A hardware update to the cori console to reduce noise on the internal electronics.2.An update to the cori system¿s software and firmware to improve the user experience when error messages are displayed, and 3.A hardware update to the cori drill to reduce mechanical stress on drill exposure the motor.The first two improvements are fully deployed.The third improvement is being deployed for new orders and as drills are returned for routine servicing.Also, smith+nephew is voluntarily performing a recall/field notification for the cori real intelligence robotic drill.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTIC DRILL
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14008785
MDR Text Key288589564
Report Number3010266064-2022-00255
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556757321
UDI-Public00885556757321
Combination Product (y/n)N
Reporter Country CodeSA
PMA/PMN Number
K201022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB10013
Device Catalogue NumberROB10013
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/10/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
CORI ROBOTICS USA- (B)(6).
Patient SexFemale
-
-