BOSTON SCIENTIFIC CORPORATION ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM; STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
|
Back to Search Results |
|
Model Number 24653 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/09/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
Device eval by manufacturer: returned product consisted of an eluvia self-expanding stent system.The outer sheath, tip, inner sheath, and the remainder of the device were checked for damage.Visual examination revealed the eluvia had a kink to the outer sheath at the nosecone.Inspection of the remainder of the device presented no other damage or irregularities.
|
|
Event Description
|
It was reported that the stent stretched with difficult deployment and an additional stent was placed to cover the area.This eluvia drug-eluting vascular stent 7mm x 120mm x 130cm was selected for use in a percutaneous transluminal angioplasty procedure using the superficial femoral artery as treatment for endovascular therapy.The 100% stenosed target lesion had moderate calcification and severe tortuosity.There was difficultly deploying the stent causing the stent to stretch, and when the post-dilation was poor they decided to deploy another eluvia 7mm x 80mm to cover fully cover the area.The procedure was considered successful, and no patient complications were reported.
|
|