According to the reporter, intra-operatively, the screen would not come on, but the lights come on.Tried batteries and other troubleshooting but did not help.There was a delay of the procedure/surgery but no harm towards the patient.
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The analysis found that there was damage to the device, but it did not interfere with the unit functioning correctly.Functionally, the laryngoscope was power cycled on and an image did not appear on the display.The light was observed from the led window.During inspection of the handle the display was found to be separated from the plastic chassis allowing for liquid ingress to occur, and a crack was found across led window.It was reported that the display was defective.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|