• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE 90-S ACCELERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE 90-S ACCELERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 0279351400
Device Problems Continuous Firing (1123); Failure to Shut Off (2939)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2022
Event Type  malfunction  
Event Description
It was reported that the device was continuously activating.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: could not be turned off.Probable root cause: "probe short, button stuck on firing position, fluid ingress, suction tube cut, button inadvertently pressed, damaged insulation, probe bender used to bend nonbendable probe, user accidentally submerges device, inadequate gluing operations (tip and core/lumen) or inadequate gluing/welding of core to handle console error: refer to rsk10684 for risk outputs associated with rf energy to probe for the serfas energy console.Refer to rsk10642 for risk outputs associated with rf signal to handpiece for crossfire and crossfire ii consoles use error.The reported failure mode will be monitored for future reoccurrence.H3 other text : 81.
 
Event Description
It was reported that the device was continuously activating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
90-S ACCELERATOR
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key14015999
MDR Text Key294488167
Report Number0002936485-2022-00163
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327057041
UDI-Public07613327057041
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K160050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0279351400
Device Catalogue Number0279351400
Device Lot Number21158AE2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2022
Initial Date FDA Received04/05/2022
Supplement Dates Manufacturer Received03/08/2022
Supplement Dates FDA Received05/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/07/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-