BOSTON SCIENTIFIC CORPORATION EPIC; STENT, ILIAC Back to Search Results
Model Number 20300
Device Problems Material Deformation (2976); Migration (4003)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/12/2022
Event Type  Injury  
Event Description
It was reported that stent damage and stent migration occurred.The 70% target lesion was located in non-tortuous and non-calcified superficial femoral artery (sfa).A 5x100x120 epic stent was deployed to treat the lesion.However, during post-dilatation, the device migrated distally and it was noted that the stent strut was compressed and misaligned in the middle portion of the stent.The lesion was missed due to migration.Furthermore, the physician decided not to removed the device in the distal sfa.Another stent was used to treat and completed the procedure.There were no patient complications nor injuries reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIC
Type of Device
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14016533
MDR Text Key288632111
Report Number2134265-2022-03186
Device Sequence Number1
Product Code NIO
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/27/2022
Device Model Number20300
Device Catalogue Number20300
Device Lot Number0024032883
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/14/2022
Initial Date FDA Received04/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-