• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA 3D IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1601-04
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Output Problem (3005)
Patient Problems Inadequate Pain Relief (2388); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 12/15/2021
Event Type  Injury  
Event Description
The recipient reportedly experienced poor performance.The recipient reportedly did not have a complete insertion of the electrode array.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed tool damage in the silicone overmold on the top cover, and cuts on the bottom cover, as well as a severed electrode.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.The device passed the mechanical test performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey mcdonald
28515 westinghouse place
valencia, CA 91355
MDR Report Key14021468
MDR Text Key288674443
Report Number3006556115-2022-00444
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016878039
UDI-Public(01)07630016878039(11)200921(17)230930
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/30/2023
Device Model NumberCI-1601-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2022
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/06/2022
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received06/02/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age47 YR
Patient SexFemale
-
-