MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ULTRATOME XL; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ULTRATOME XL; SNARE, FLEXIBLE Back to Search Results
Model Number M00535900
Device Problems Break (1069); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/11/2022
Event Type  malfunction  
Event Description
It was reported to boston scientific corporation that an ultratome xl was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the physician inserted the device through the duodenoscope to perform sphincterotomy.When the device was energized, the "wire burst." it did not cause any damage to the patient.Additionally, no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an ultratome xl was used in the papilla during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6), 2022.During the procedure, the physician inserted the device through the duodenoscope to perform sphincterotomy.When the device was energized, the "wire burst." it did not cause any damage to the patient.Additionally, no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Note: a photo of the complaint device outside the patient was provided by the customer and shows the cutting wire was broken and kinked.
 
Manufacturer Narrative
Block h6 (device codes): medical device problem code a0401 captures the reportable event of cutting wire broken.Block h10: the returned ultratome xl was analyzed, and a visual evaluation noted that the cutting wire was broken and bent from the proximal pierced hole.The device was observed under magnification and the cutting wire was broken, bent, and blackened from the proximal pierced hole.These findings were consistent with the finding per media analysis based on the provided photo.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, and blackened.The cutting wire being blackened indicates that the device was energized.Based on the condition of the device, the problem found could have been caused if the cutting wire was preactivated before use which can cause premature cutting wire fatigue and compromise the cutting wire's integrity.Also, not verifying that the cutting wire has exited the endoscope when electrical current is applied and not maintaining direct and constant contact with tissue when applying electrocautery current could contribute to wire breakage.Additionally, an excess of power could have damaged the integrity of cutting wire.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This can kink/bend the cutting wire.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATOME XL
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key14023559
MDR Text Key288700135
Report Number3005099803-2022-01820
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729103257
UDI-Public08714729103257
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K930022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/02/2022
Device Model NumberM00535900
Device Catalogue Number3590
Device Lot Number0026121884
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/31/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/02/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-