• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE AFFINITY 4 BED FRAME; TABLE, OBSTETRIC (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE AFFINITY 4 BED FRAME; TABLE, OBSTETRIC (AND ACCESSORIES) Back to Search Results
Model Number P3700B000107
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/23/2022
Event Type  malfunction  
Manufacturer Narrative
The hillrom service technician identified that the brakes would set but would not hold.The technician replaced the brake detent mechanism assembly to resolve this issue.The affinity® 4 birthing bed is intended to be used as a birthing bed for women of child bearing age in an ldr (labor, delivery,recovery) or ldrp (labor, delivery, recovery, postpartum) setting within the acute care labor and delivery market.It is not intended for use as a general hospital bed.The affinity iv brake and steer pedals are connected in the center of the base frame by the detent assembly which is used to synchronize the action of the brake/steer functions from either the left or right sides of the bed.When the pedal is pressed down into the brake position, this causes the detent assembly to actuate the brake linkage of all four casters, therefore placing all four casters into brake.When the pedal is pressed down into the steer position, the brake linkage will only actuate the foot left brake, thus leaving the other three casters in a free swivel condition.When the detent is broken the brake and steer functions will not operate properly, thus allowing the brake/steer functionality to be uncontrolled.A core component of the affinity iv braking system is the brake detent assembly.The brake detent assembly for this affinity iv bed is comprised of a plastic injection molded housing, steel detent coupler with pins, steel plunger and a compression spring.Hillrom¿s investigation results indicate the most probable root cause of the malfunction of the detent housing is tensile stress combined with wear from use.Frequent brake/neutral/steer activations during use of the bed may contribute to bed¿s brake system wear.If the brake detent housing malfunctions, the user is immediately aware as the braking system will not engage on the bed.The global complaint rate continues to be low for detent malfunctions for 2021.There have been no reports of injuries or deaths related to this failure mode.Hillrom considers this to be a highly detectable failure mode since as it occurs the user will recognize the brakes are not engaging and remove the bed from service until repaired.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in (b)(6) 2021.It is unknown if the facility performed any other preventative maintenance on this bed.Additionally, per the hillrom's service manual, the affinity® three birthing bed and affinity® four birthing beds require an effective maintenance program.A semiannual preventive maintenance schedule is recommended which includes checking the brakes.The manual instructs: apply the brake, and check to ensure that the bed will not move.If the bed moves, inspect it for wear, and adjust if required.Apply the steering pedal and check the steering to ensure proper locking action when activated.Based on this information, no further action is required.
 
Event Description
Hillrom received a report from a hillrom technician stating the bed brakes were not holding.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFINITY 4 BED FRAME
Type of Device
TABLE, OBSTETRIC (AND ACCESSORIES)
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
kayla miller
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key14045094
MDR Text Key291505818
Report Number1824206-2022-00173
Device Sequence Number1
Product Code KNC
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K915779
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP3700B000107
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/23/2022
Date Device Manufactured02/26/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-