• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PACEL BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PACEL BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number 401771
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/03/2022
Event Type  malfunction  
Event Description
The patient was admitted due to myocardial ischemia caused by a myocardial infarction.During hospitalization, the monitoring showed slow heart rate, as low as 30 bpm.On (b)(6) 2020, the patient underwent temporary pacemaker implantation, and the heart rate was 80-90 beats/min when the temporary pacemaker was used.There was no abnormality during the surgery.The ecg monitoring showed that the heart rate was 45 bpm.The patient was given isoproterenol to increase the heart rate.The ecg monitoring showed that the heart rate was 60 bpm and the temporary pacemaker was not needed.On the evening of the second postoperative day, the patient's family helped the patient turn over the slightly and found that the temporary pacemaker fell off.The patient was in currently in stable condition.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported detachment could not be conclusively determined.
 
Event Description
The temporary pacemaker detached where the device was sutured to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACEL BIPOLAR PACING CATHETER
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345 2126
Manufacturer (Section G)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345 2126
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14046131
MDR Text Key288866748
Report Number2182269-2022-00022
Device Sequence Number1
Product Code LDF
UDI-Device Identifier05414734001137
UDI-Public05414734001137
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K152784
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Model Number401771
Device Catalogue Number401771
Device Lot Number7476166
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/08/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-