Model Number MODEL 100
Device Problems Display Difficult to Read (1181); Display or Visual Feedback Problem (1184)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/10/2022
Event Type  malfunction  
Event Description
During shift check, the customer reported the lcd display of the autopulse platform(sn (b)(4)) is flashing and not readable.Also, the brake intermittently powers on and off.No patient involvement.
 
Manufacturer Narrative
The autopulse platform involved in the reported complaint will not be returned to a zoll service depot for evaluation.The device was investigated and serviced by zoll trained distributor in brazil.The defective processor board (pca) was replaced to remedy the unreadable lcd screen and intermittent brake issue.Also, the load plate cover was replaced.Since the issues were resolved, service was not required to be performed by zoll.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer (Section G)
ZOLLCIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave
san jose, CA 95131
4084192922
MDR Report Key14046209
MDR Text Key288831435
Report Number3010617000-2022-00334
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000031
UDI-Public00849111000031
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-24
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/10/2022
Initial Date FDA Received04/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/30/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-