Model Number G148
Device Problem Failure to Convert Rhythm (1540)
Patient Problems Arrhythmia (1721); Cardiac Arrest (1762); Electric Shock (2554)
Event Date 01/26/2022
Event Type  Injury  
Manufacturer Narrative
This investigation will be updated should further pertinent information be provided.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) showed an episode were a ventricular arrhythmia was not successfully converted.The patient was taken back after a defibrillation threshold (dft) test that was not successful.The crt-d was changed from right side to left side of the body and the rv and ra leads were explanted without allegations against the leads.The patient was dependent, therefore experienced a cardiac arrest outside the hospital.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14048508
MDR Text Key288857275
Report Number2124215-2022-08213
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/29/2023
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number261869
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/26/2022
Initial Date FDA Received04/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/16/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexMale
-
-