It was reported that there was an issue with ff904sb - caspar distraction pin 14mm.According to the complaint description, during the removal of the distraction pins, the head of 2 pins broke.Portions of the threads were left in the patient's body at c5 and c6.This occurred during a neurosurgical procedure.A fragment remained in situ.Additional information was not provided.Additional patient information is not available.The adverse event is filed under aag reference (b)(4).
|
Updated d9: not returned to manufacturer for evaluation.Updated h6: codes.Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
|