• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC OPEN SPINE CLAMP; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC OPEN SPINE CLAMP; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9731780
Device Problem Mechanics Altered (2984)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2022
Event Type  malfunction  
Event Description
Medtronic received information regarding a navigation system being used outside of a procedure.The spine clamp was damaged.There was fixation failure due to abrasion at the spike region.There was no patient involvement.
 
Manufacturer Narrative
Analysis found on the clamp- analysis determined the returned clamp has a few bent or blunted tips on the spiked surfaces of the two clamp faces.The clamp is otherwise fully functional.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN SPINE CLAMP
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14053618
MDR Text Key288896785
Report Number1723170-2022-00523
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K990214
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9731780
Device Catalogue Number9731780
Device Lot Number160818
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2022
Initial Date Manufacturer Received 03/28/2022
Initial Date FDA Received04/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-