BOSTON SCIENTIFIC CORPORATION EASYTRAK 2; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D Back to Search Results
Model Number 4517
Device Problem High impedance (1291)
Patient Problems Fall (1848); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2022
Event Type  Injury  
Event Description
It was reported that this left ventricular (lv) lead exhibited a high out of range pacing impedance measurements of greater than 2000 ohms.This measurement coincided with a time in which the patient had reported experiencing a fall.No changes have been made and the lv lead remains in service.No adverse patient effects were reported.
 
Event Description
It was reported that this left ventricular (lv) lead exhibited a high out of range pacing impedance measurements of greater than 2000 ohms.This measurement coincided with a time in which the patient had reported experiencing a fall.No changes have been made and the lv lead remains in service.No adverse patient effects were reported.
 
Manufacturer Narrative
As no further information regarding this event is expected, our investigation is complete.This report will be updated should additional information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
CPI - DEL CARIBE
guidant puerto rico b. v.
#12, rd. #698
dorado PR 00646
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14068821
MDR Text Key288991821
Report Number2124215-2022-08191
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526417665
UDI-Public00802526417665
Combination Product (y/n)Y
Reporter Country CodeIT
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/27/2005
Device Model Number4517
Device Catalogue Number4517
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/11/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/27/2003
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-