OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problem Noise, Audible (3273)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Ne device was received for evaluation.Visual inspection found normal wear and tear of the enclosure and front cover.The tank cover was cracked and the pole clamp was broken.The pcb and power switch were outdated.The tank was filled with water, the temp check attached and the device was plugged in.Functional testing confirmed the pump was noisy.Due to the age and condition of the device the root cause is not attributed to its manufacture.
 
Event Description
It was reported that the pump was very noisy.No patient involvement or injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14075397
MDR Text Key289019042
Report Number3012307300-2022-06171
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2022
Was the Report Sent to FDA? No
Date Manufacturer Received03/16/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/26/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-