Model Number 0998-00-3023-55 |
Device Problems
Premature Discharge of Battery (1057); Failure to Run on Battery (1466)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/14/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) failed the battery run time test.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
|
|
Manufacturer Narrative
|
Type of investigation not yet determined: (4118/3233): a supplemental report will be submitted upon receipt of additional information.
|
|
Manufacturer Narrative
|
Corrected field: medical device- problem code.A getinge field service engineer (fse) replaced the batteries to resolve the issue.The unit passed all functional and safety tests to meet manufacturer specifications.The iabp was returned to the customer and cleared for clinical use.The initial reporter named in block e1 is a getinge employee who has different contact details from that of the event site.A contact person at the event site is katie malcolmson, katie.Malcolmson@bmihealthcare.Co.Uk.
|
|
Event Description
|
During preventative maintenance (pm) conducted by a getinge field service engineer (fse) the cs300 intra-aortic balloon pump (iabp) failed the battery run time test.There was no patient involvement and no adverse event reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Additional information: event site postal code - (b)(6).
|
|