• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-55
Device Problems Premature Discharge of Battery (1057); Failure to Run on Battery (1466)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2022
Event Type  malfunction  
Event Description
It was reported that the cardiosave intra-aortic balloon pump (iabp) failed the battery run time test.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Manufacturer Narrative
Type of investigation not yet determined: (4118/3233): a supplemental report will be submitted upon receipt of additional information.
 
Manufacturer Narrative
Corrected field: medical device- problem code.A getinge field service engineer (fse) replaced the batteries to resolve the issue.The unit passed all functional and safety tests to meet manufacturer specifications.The iabp was returned to the customer and cleared for clinical use.The initial reporter named in block e1 is a getinge employee who has different contact details from that of the event site.A contact person at the event site is katie malcolmson, katie.Malcolmson@bmihealthcare.Co.Uk.
 
Event Description
During preventative maintenance (pm) conducted by a getinge field service engineer (fse) the cs300 intra-aortic balloon pump (iabp) failed the battery run time test.There was no patient involvement and no adverse event reported.
 
Event Description
N/a.
 
Manufacturer Narrative
Additional information: event site postal code - (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key14076172
MDR Text Key289776593
Report Number2249723-2022-00792
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567112541
UDI-Public10607567112541
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-55
Device Catalogue Number0998-00-3023-55
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/24/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2013
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.; UNKNOWN.
-
-