Udi: (b)(4).No product was returned to nova biomedical for investigation.No manufacturer retains of the test strip lot: 0320070249 were available for testing due to the age of the consumable which was already expired at the time nova biomedical was made aware of the alleged malfunction.The similar lot: 0320153249 was tested in place of the reported lot as this was manufactured only a few months later.The alternate test strips were tested using five retained meters during analysis of multiple samples.During this usage, the reported error was not seen with the retained strips and the reported problem could not be reproduced as all used test strips and meters were able to produce accurate results after applying the sample.Device history record (dhr) reviews were performed for the test strips being used at the time of the event by a quality control investigator.The reviews included an assessment of the production, testing, and release of the meter and test strips.No abnormalities or concerns were noted, and the dhr indicated the released product met all specifications.The conclusion of the investigation is the reported customer complaint could not be reproduced after testing retains of similar aged test strips.A conclusive root cause was unable to be identified, and nova will continue to monitor for recurrence of similar events.
|