TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS TMJ BILATERAL IMPLANTS; PATIENT-FITTED TMJ RECONSTRUCTION PROSTHESIS Back to Search Results
Catalog Number CHG020
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 04/01/2022
Event Type  Injury  
Manufacturer Narrative
Device is not available for evaluation.If additional information is received it will be reported on a supplemental report.
 
Event Description
It was reported that the patient had developed heterotopic bone.The surgeon plans to remove the bilateral devices, resect additional mandibles and replace new devices in one stage additional surgery.
 
Event Description
It was reported that the patient had developed heterotopic bone.The surgeon plans to remove the bilateral devices, resect additional mandibles and replace new devices in one stage additional surgery.
 
Manufacturer Narrative
This event could be confirmed based on the ct imaging received.A new order was submitted on stryker¿s id portal, and a new bilateral set of devices are being designed, manufactured, and shipped to the hospital.Heterotopic bone is a patient driven failure mode and unrelated to the tmj device itself.This event does not involve any device failure, malfunction, or performance issues.Based on the investigation, there is no indication of an incorrectly working product or any design, material, or manufacturing-related issues.Therefore, no corrective and/or preventive actions are deemed necessary at this time.The complaint is added to the complaint trend.H3 other text : not available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TMJ BILATERAL IMPLANTS
Type of Device
PATIENT-FITTED TMJ RECONSTRUCTION PROSTHESIS
Manufacturer (Section D)
TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS
6059 king drive
ventura 93003
Manufacturer (Section G)
TMJ SOLUTIONS, LLC DBA TMJ CONCEPTS
6059 king drive
ventura 93003
Manufacturer Contact
rylee koole
6059 king drive
ventura 93003
8056503391
MDR Report Key14078693
MDR Text Key289060052
Report Number0002031049-2022-00018
Device Sequence Number1
Product Code LZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980052
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberCHG020
Device Lot NumberT17-0121
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/01/2022
Initial Date FDA Received04/11/2022
Supplement Dates Manufacturer Received04/01/2022
Supplement Dates FDA Received07/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/17/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age58 YR
-
-