MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problem Problem with Software Installation (3013)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2021
Event Type  malfunction  
Event Description
It was reported that the programmer which was returned for service subsequently tested out of specification during manufacturer's assessment.There was no patient involvement.
 
Manufacturer Narrative
Product analysis: it was determined on analysis that the programmer tested out of specification as the stylus tip position on the touch screen had a s light deviation and required calibration.It was also noted that the stylus was missing.The key board connection cable shielding was damaged.The upper and lower bullet assembly was broken.The printer paper tray was empty.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key14082517
MDR Text Key289057514
Report Number2182208-2022-01131
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2022
Date Manufacturer Received03/30/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/27/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-