Model Number 0998-00-0800-55 |
Device Problem
Failure to Run on Battery (1466)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/16/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the cardiosave intra-aortic balloon pump (iabp) did not work on battery when the power plug was disconnected.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
|
|
Manufacturer Narrative
|
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to evaluate the iabp and was confirmed the reported issue.The fse determined that although the batteries are full, it did not switch to work from the battery when the mains voltage was cut.To fix the issue, the fse replaced the power supply (0670-00-1162), and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that during a routine check, the cardiosave intra-aortic balloon pump (iabp) did not work on battery when the power plug was disconnected.There was no patient involved and no adverse event was reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
H6 (component codes and health impact code).
|
|