• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Model Number P310NUS
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/15/2022
Event Type  malfunction  
Event Description
It was reported that the flow switch is not working and cpap flow will not turn on.No patient involvement was reported.
 
Manufacturer Narrative
Other text: h6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device was received in with a cracked peep knob and missing four small end caps, the cpap knob contains marks, bottom casing contains a minor crack, the threads on the vrv block manifold have stripped, variable relief valve is now loose.Internal inspection shows contamination of foreign material.The customer stated problem was duplicated, unable to spin the knob to its min and max settings.Replaced bottom chassis, vrv block manifold, housing, function switch, and 2 small knobs.Installed mri battery, sintered filter, and o-rings for the pm.Reset patient pressure cycling led and adjusted cpap control to tolerance.Performed pm, cleaned device and affixed new service label.The cause of the reported problem was traced to the customer manipulation of the device.The previous service history has been reviewed and is deemed unrelated to the current customer reported problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC PLUS
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK  LU3 4BU
Manufacturer (Section G)
NULL
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK   LU3 4BU
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14089849
MDR Text Key289174623
Report Number3012307300-2022-06245
Device Sequence Number1
Product Code BTL
UDI-Device Identifier35019315107263
UDI-Public35019315107263
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP310NUS
Device Catalogue NumberP310NUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/18/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received05/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/18/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-