• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Catalog Number CON-HL-90
Device Problems Mechanical Problem (1384); Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that there was over temping and water not circulating.No patient involvement was reported.
 
Manufacturer Narrative
One device was received and evaluated.Visual inspection found the pump was faulty, wear and tear of enclosure and front cover and a faded line cord.Functional testing found the overtemp alarm was operating correctly.The pump was observed to not be circulating water.The faulty pump caused the water not circulating issues.The root cause can not be attributed to manufacturing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key14101742
MDR Text Key289200357
Report Number3012307300-2022-06290
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/17/2022
Initial Date FDA Received04/13/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/24/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-