DEPUY IRELAND - 9616671 ATTUNE DISTAL FEMORAL JIG; ATTUNE INSTRUMENTS : ALIGNMENT DEVICES Back to Search Results
Model Number 2544-00-520
Device Problems Break (1069); Structural Problem (2506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/28/2022
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The attune distal femoral jig is damaged.The spring on the small attachment has failed.No surgical delay.No adverse affects on patient.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> the device associated with this report was returned for examination.Visual examination of the returned device and photo evidence reviewed exhibited that the spring of the slide tube disassembled, making the device unable to lock on mating device.The reported condition was confirmed.There was not observed any breakage condition at the device.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot
=
> the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE DISTAL FEMORAL JIG
Type of Device
ATTUNE INSTRUMENTS : ALIGNMENT DEVICES
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14102652
MDR Text Key289185073
Report Number1818910-2022-06663
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295434030
UDI-Public10603295434030
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2544-00-520
Device Catalogue Number254400520
Device Lot NumberABC50471
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/29/2022
Initial Date FDA Received04/13/2022
Supplement Dates Manufacturer Received04/20/2022
06/20/2022
Supplement Dates FDA Received04/20/2022
06/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-