Model Number 0475100000
Device Problems Thermal Decomposition of Device (1071); Circuit Failure (1089)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2022
Event Type  malfunction  
Event Description
It was reported that there was a thermal event.
 
Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event Description
It was reported that there was a thermal event.
 
Manufacturer Narrative
Alleged failure: smoke and burning smell confirmed failure: csl-r-pcb circuit board failure probable root cause: console fuses fail to break circuit due to excessive mains current leading to fire isolation power supply failure use error: console is sterilized or disinfected use error: console cleaned with incompatible products the product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, CROSSFIRE 2 CONSOLE
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
andrea zenere
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key14103054
MDR Text Key294777337
Report Number0002936485-2022-00178
Device Sequence Number1
Product Code GEI
UDI-Device Identifier07613327058109
UDI-Public07613327058109
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071859
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0475100000
Device Catalogue Number0475100000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received03/17/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-