Catalog Number UNK HIP FEMORAL HEAD METAL |
Device Problem
Naturally Worn (2988)
|
Patient Problems
Adhesion(s) (1695); Hypersensitivity/Allergic reaction (1907); Pain (1994); Loss of Range of Motion (2032); Scar Tissue (2060); Distress (2329); Discomfort (2330); Metal Related Pathology (4530); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 11/03/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).Initial reporter occupation: (b)(6).(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
Pinnacle mom litigation record received.Patient alleges heavy metal poisoning from the toxic heavy metals, pain, suffered additional scar tissue formation, suffered injury to his muscle and tissue, and now has a hip replacement with decreased longevity.As a direct and proximate result of the defective pinnacle hip replacement, plaintiff suffered injuries, including but not limited to tissue destruction, muscle destruction, pseudotumor, metal wear, metal poisoning, loss of enjoyment of life, emotional distress, and limitation of daily activities.Doi: (b)(6) 2008.Dor: (b)(6) 2021.(unk hip).
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre) will not be performed since mom systems are obsolete.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthese joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
The medical record received indicates that on (b)(6) 2021, the patient was revised due to failed right hip replacement with metal-on-metal pseudotumor.The patient experienced increasing pain and discomfort on the right hip and reported elevated cobalt and chromium levels.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthese joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthese joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: a4, b5, b6, b7.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Medical records were received: after review of medical records patient was revised due to failed right hip replacement with metal on metal pseudotumor.Operative notes indicated increasing pain and discomfort, a large pseudotumor mass directly over his greater trochanter.This was slowly peeled off the greater trochanter.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H6 clinical code: metal related pathology (e1618) reported in the previous medwatch was used to capture metal poisoning and blood heavy metal increased.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Medical records received.Clinic visit complaints and vascular labs as the replacement device was polyethylene.On (b)(6) 2022 right leg pain, swelling, knee swelling, us venous duplex showed no sonographic evidence of dvt.In the medical records provided, there was no allegation of invasive procedure or revision.
|
|
Search Alerts/Recalls
|