• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. UNKN FEMORAL HEAD IMPL; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. UNKN FEMORAL HEAD IMPL; PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL, Back to Search Results
Catalog Number UNKNOWN
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Failure of Implant (1924)
Event Date 03/22/2022
Event Type  Injury  
Event Description
It was reported that a revision surgery was performed due to liner wear.The surgeon says the devices were not defective.The current health status of patient it is unknown.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
H3, h6: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, smith and nephew has not received the device/ adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Potential causes could include but are not limited to friction, joint tightness or lifetime of device.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN FEMORAL HEAD IMPL
Type of Device
PROSTHESIS, HIP, HEMI-, TRUNNION-BEARING, FEMORAL,
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14104612
MDR Text Key289207944
Report Number1020279-2022-01818
Device Sequence Number1
Product Code JDH
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2022
Initial Date FDA Received04/13/2022
Supplement Dates Manufacturer Received05/19/2022
Supplement Dates FDA Received05/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-