This report is 3 of 4 for (b)(4).If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: visual analysis of the photo revealed that drill bit ø2.5 l110/85 2flute f/qc was observed with some scratched and the tip appears to be dull.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|