• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH DRILL BIT Ø2.5 L110/85 2FLUTE F/QC; INSTRUMENT, SURGICAL, ORTHOPEDIC, PNEUMATIC POWERED & ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH DRILL BIT Ø2.5 L110/85 2FLUTE F/QC; INSTRUMENT, SURGICAL, ORTHOPEDIC, PNEUMATIC POWERED & ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 310.250
Device Problem Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2022
Event Type  malfunction  
Manufacturer Narrative
This report is 3 of 4 for (b)(4).If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: visual analysis of the photo revealed that drill bit ø2.5 l110/85 2flute f/qc was observed with some scratched and the tip appears to be dull.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Meter tip is bent.The stardrive is damaged and the bits are dull.The patient's bone was burned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL BIT Ø2.5 L110/85 2FLUTE F/QC
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, PNEUMATIC POWERED & ACCESSORY/ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14115354
MDR Text Key289328682
Report Number8030965-2022-02439
Device Sequence Number1
Product Code HSZ
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number310.250
Device Lot Number107P272
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/16/2022
Initial Date FDA Received04/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
DEPTH GAUGE F/SCR Ø4.5-6.5 MEAS-RANGE UP; DRILL BIT Ø2 W/DOUBLE MARKING L140/115 3; DRILL BIT Ø2 W/DOUBLE MARKING L140/115 3; LOCKSCR Ø5 SELF-TAP L24 SST
-
-