ST. JUDE MEDICAL, INC.(CRM-SYLMAR) MERLIN¿; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 3650
Device Problem Program or Algorithm Execution Failure (4036)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/24/2022
Event Type  malfunction  
Event Description
It was reported the device was subject to the decrement test field action issued by abbott on 10 march 2022.No patient consequences.
 
Manufacturer Narrative
The reported event of a delay during cancellation of capture test was confirmed.Based on the information provided, it was determined that the delay was a result of a software anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN¿
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14119046
MDR Text Key289342858
Report Number2017865-2022-08083
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734500661
UDI-Public05414734500661
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial,Followup
Report Date 12/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Device Catalogue Number3650
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2017865-03/10/22¿001¿C
Patient Sequence Number1
Treatment
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR; RIGHT ATRIAL LEAD; RIGHT VENTRICLE LEAD
Patient Age73 YR
-
-