Model Number 6662 |
Device Problem
Wireless Communication Problem (3283)
|
Patient Problem
Inadequate Pain Relief (2388)
|
Event Date 03/28/2022 |
Event Type
Injury
|
Event Description
|
It was reported that following an unrelated surgical procedure wherein the ipg was not placed into surgery mode, the ipg became unable to communicate with external devices.Troubleshooting was unable to resolve the issue.As a result, surgical intervention was undertaken on (b)(6) 2022 wherein the ipg was explanted and replaced to address the issue.
|
|
Manufacturer Narrative
|
Date of event is estimated.The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.
|
|
Manufacturer Narrative
|
An inoperable implant was reported to abbott.The system was not set to surgery mode while the patient underwent an unrelated surgery where electro-cautery may have been used.The implant was not returned for analysis.A review of documentation supplied with the implant states that electro-surgery devices should not be used in close proximity to an implanted system.Based on the information received, the cause of the reported incident is consistent with user error.
|
|