• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 DXTEND SCREW LOCK D4.5X36MM; DELTA XTEND IMPLANTS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 DXTEND SCREW LOCK D4.5X36MM; DELTA XTEND IMPLANTS : SCREWS Back to Search Results
Model Number 130790036
Device Problem Osseointegration Problem (3003)
Patient Problems Inadequate Osseointegration (2646); Joint Laxity (4526)
Event Date 03/30/2022
Event Type  Injury  
Event Description
Revision completed due to loosening of the metaglene at the bone to implant interface.It was also reported that the 12 year old delta reverse became unstable, poly wear is present, and this is a glenoid defect.Patient converted into a cta, stem was well fixed.Doi- unknown.Dor- (b)(6) 2022.Affected side- left shoulder.
 
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X36MM
Type of Device
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI  
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14136996
MDR Text Key289454273
Report Number1818910-2022-06856
Device Sequence Number1
Product Code PHX
UDI-Device Identifier10603295027850
UDI-Public10603295027850
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K120174
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130790036
Device Catalogue Number130790036
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DXTEND GLENOSPHERE ECC D42MM; DXTEND METAGLENE; DXTEND SCREW LOCK D4.5X24MM; DXTEND SCREW LOCK D4.5X42MM; DXTEND SCREW NO LOCK D4.5X18MM; DXTEND STAND PE CUP D42 +6MM; UNK SHOULDER HUMERAL EPIPHYSIS DELTA XTEND
Patient Outcome(s) Required Intervention;
Patient Age74 YR
Patient SexMale
-
-