• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESMED LTD ASTRAL BATTERY PACK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESMED LTD ASTRAL BATTERY PACK Back to Search Results
Model Number 19620
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/20/2022
Event Type  malfunction  
Event Description
It was reported to resmed that an astral device displayed an error message (sf180) related to a battery charger fault.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The device was returned to an authorized resmed third party service center and an evaluation confirmed the complaint.The internal battery was replaced to address this issue.The device was serviced and fully tested before it was returned to the customer.Resmed reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL BATTERY PACK
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
Manufacturer (Section G)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
Manufacturer Contact
michelle hughes
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS   2153
MDR Report Key14137581
MDR Text Key294798734
Report Number3004604967-2022-00255
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K152068
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation 505
Type of Report Initial
Report Date 04/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number19620
Device Catalogue Number19620
Was Device Available for Evaluation? No
Distributor Facility Aware Date03/20/2022
Initial Date Manufacturer Received 03/20/2022
Initial Date FDA Received04/18/2022
Date Device Manufactured09/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-