Catalog Number UNK SUPERA |
Device Problem
Migration (4003)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/12/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Estimated date of event.The udi number is not known as the part and lot number were not provided.Estimated date of implant.The device was not returned for analysis.A review of the lot history record and similar incident review of the reported lot could not be conducted because the part and lot number was not provided.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that when trying to insert the introducer in the cfa interaction with the previously implanted supera stent resulted in the reported migration of the previously implanted supera stent.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|
Event Description
|
It was reported that the procedure was to treat the common femoral artery(cfa).When trying to insert the introducer in the cfa before the procedure, the introducer moved the supera stent which was already implanted in the cfa, further down.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
|