Additional information: h3, h6, h10.H10: the device was received for evaluation.A visual inspection with the naked eye and magnification noted a broken occlude foot.No issues were noted during passage testing.During leak and clamp function testing the set leaked with the clamp in closed position.The reported condition was verified.The cause of the condition could not be determined; however, the damaged set is consistent with damage caused by exposure to chemical agents such as foreign solvents, dyes, or cleaning agents that damage the transfer set materials.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|