DATEX-OHMEDA, INC. CARESTATION 650; ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1012-9650-000
Device Problem Gas/Air Leak (2946)
Patient Problem Low Oxygen Saturation (2477)
Event Date 03/21/2022
Event Type  Injury  
Event Description
The hospital reported a patient was connected to a carestation 650 when the unit was suspected to leak oxygen during ventilation.Reportedly, the patient desaturated.Oxygen was increased and patient oxygen saturation level returned to normal.There was no patient injury.
 
Manufacturer Narrative
Ge healthcare's investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.No patient information provided to date.Unique identifier: (b)(4).Legal manufacturer: (b)(4).Device evaluation anticipated, but not yet begun.
 
Manufacturer Narrative
Leakage potentially occurred from the device or the breathing circuit.However, as no failure with the device was found and the customer did not have access to the breathing circuit, the leakage source could not be identified.Based on available information, the root cause of the patient desaturation was undetermined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key14142974
MDR Text Key289520257
Report Number2112667-2022-00926
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1012-9650-000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-