• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR DRAGONFLY OPSTAR¿ IMAGING CATHETER; DIAGNOSTIC IMAGING CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR DRAGONFLY OPSTAR¿ IMAGING CATHETER; DIAGNOSTIC IMAGING CATHETER Back to Search Results
Catalog Number 1014652
Device Problem Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The bmw guide wire is filed under separate medwatch report number.
 
Event Description
It was reported the procedure was performed in the right coronary artery.A bmw guide wire was inserted.Predilation was performed, then a dragonfly opstar catheter was inserted.However, while attempting to remove the catheter, it was observed the guide seemed blocked and was unable to be removed.Therefore, the catheter and the guide wire were removed all as one.A new wire was used to continue the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
Visual and dimensional analysis was performed on the returned devices.The reported difficulty removing the catheter was unable to be confirmed due to the device condition; however, damage to the dragonfly catheter which is consistent with the reported event was observed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.A query of the complaint handling database for the reported lot was performed and revealed there is no indication of a lot level quality product issue.The investigation determined that the difficulties were likely related to circumstances of the procedure.The torn guide wire exit port noted on the returned dragonfly catheter suggests that the guidewire and catheter were likely spread apart during removal causing the guide wire to tear into the exit port of the dragonfly resulting in difficulty removing and damage to the guide wire tip coils.While the reported withdrawal issue could not be confirmed, it is likely that either the patient¿s anatomical condition(s) or the guide catheter size affected the withdrawal; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY OPSTAR¿ IMAGING CATHETER
Type of Device
DIAGNOSTIC IMAGING CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT MEDICAL REG#3009600098
4 robbins rd
westford MA 01886
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14145893
MDR Text Key289600743
Report Number2024168-2022-04219
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K192019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/19/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/18/2023
Device Catalogue Number1014652
Device Lot Number8211511
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/18/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BMW GUIDE WIRE
-
-